Numerical study of the process of optical radiation measuring using a hollow aperture cantilever

Read full text
Authors: Elena Sergeevna Kozlova, Sergey S. Stafeev, Victor Victorovich Kotlyar
Cited by: 0
References: 44

Abstract

Background. Scanning near-field optical microscopy is a powerful and promising tool for studying the interaction of light with matter. Improving the quality and accuracy of measurements is a key challenge. When interpreting experimental results, it is important to consider the sensitivity of the cantilever to the polarization type. Aim. The aim of this study is to evaluate the sensitivity of a pyramidal aperture cantilever to the incidence angle of measured light. Methods. Using the frequency depended finite-difference time-domain method, it was shown that the incidence angle and the plane of incidence can affect the measurement process with an aperture cantilever for both proposed types of incident light polarization: linear and circular left-handed. Results. Simulation results show that with an increase in the incidence angle of measured light, the overall intensity inside the cantilever decreases by approximately 50 % and 30 % for linearly and circularly polarized light, respectively. Conclusion. This proves that the aluminum aperture cantilever is weakly sensitive to the longitudinal component.

Fields of Study & Subjects

Near-Field Optical Microscopy Plasmonic and Surface Plasmon Research Microwave and Dielectric Measurement Techniques

Publication details

Journal / source
Physics of Wave Processes and Radio Systems
Volume
28
Issue
4
Pages
97–109
Publication date
27 January 2026
Language
RU