Establishing Pediatric Thoracic Radiography Diagnostic Reference Levels Using CALDOSE_X: A Data-Driven Approach to Optimize Radiation Safety

Read full text
Authors: I. Bouadel, Morad El Kafhali, Marziyeh Tahmasbi, Soukaina Mefrah, Rajaa Sebihi
Cited by: 1

Abstract

Purpose: Disparities exist in adherence to national radiation safety standards in Morocco, particularly in pediatric conventional radiology. This cross-sectional study aims to establish Moroccan diagnostic reference levels (DRLs) for pediatric thorax radiography. Materials and Methods: Thorax radiographs of 208 pediatric patients (newborns to 18 years old) from four Moroccan public hospitals were analyzed. Patient demographics (age, gender, weight) and scan parameters were recorded to calculate radiation doses using CALDOSE_X 5.0 software, focusing on entrance surface air kerma (ESAK, mGy) and kerma-area product (KAP, mGy·cm²). Patients were categorized into five age groups (

Fields of Study & Subjects

Radiation Dose and Imaging Digital Radiography and Breast Imaging Advanced Radiotherapy Techniques

Publication details

Journal / source
Frontiers in biomedical technologies
Publisher
Knowledge E DMCC
Publication date
27 January 2026