The Safety and Efficacy of Disinfection Cap Use in a Real-World Environment
Abstract
Purpose: Catheter hub and needle-free connector contamination is a causative element in the development of catheter-related bloodstream infections and demonstrates the necessity for effective hub disinfection. The current study evaluated the safety and efficacy of a disinfection cap used in a real-world environment. Patients and Methods: This multicenter, single-arm, prospective post-market study was designed to collect clinical data on the safety and performance of a passive disinfection cap. A total of 157 participants were included in this study. The primary performance endpoints were attachment and removal success rates. The primary safety endpoint was the rate of device-related AEs. Results: A total of 156 participants, aged 3-93 years, had at least one disinfection cap applied during the study. Attachment and removal success was ≥ 99.7%. No device-related adverse events were observed. Conclusion: This study found that disinfection caps can be attached and removed successfully with no device-related adverse events in a real-world environment.
Fields of Study & Subjects
Central Venous Catheters and Hemodialysis
Infection Control in Healthcare
Urinary Tract Infections Management
Publication details
- Journal / source
- Medical Devices Evidence and Research
- Volume
- Volume 19
- Pages
- 1–8
- Publication date
- 1 February 2026
- Language
- EN