Attaining the optimal performance of the CMS lead tungstate Electromagnetic Calorimeter in the Run3 of LHC

Read full text
Authors: P. Rebello Teles
Cited by: 0

Abstract

The CMS Electromagnetic Calorimeter (ECAL) plays a central role in precision measurements involving electrons, photons, and jets. During Run3 of the LHC, increased luminosity and detector ageing posed significant challenges for maintaining optimal energy resolution and stability. Since 2024, ECAL calibration workflows have been fully automated, resulting in a tenfold reduction in calibration latency. This paper presents the methods and automation framework, and shows results demonstrating stable ECAL energy resolution of about 2% in the central barrel and better than 5% in the endcaps, with a stable energy scale across years of data taking.

Fields of Study & Subjects

Particle Detector Development and Performance Particle physics theoretical and experimental studies Superconducting Materials and Applications

Publication details

Pages
526–526
Publication date
6 January 2026