Accuracy of Continuous Glucose Monitoring in Adults With Post-Bariatric Hypoglycemia After Roux-en-Y Gastric Bypass Surgery Under Stable and Dynamic Glucose Conditions

Read full text Summarise with VIRAT
Authors: Aritz Lizoain, Martina Vettoretti, Antonio Ferreira, Katja A. Schönenberger, Olivia Streicher, Eva Rolfes, Vincent Braunack-Mayer, Andrea Facchinetti, David Herzig, Martina Rothenbühler, Lia Bally
Cited by: 0

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

Objective: Although continuous glucose monitors (CGMs) are increasingly used to detect and manage post-bariatric hypoglycemia (PBH) and associated glucose variability, data on their accuracy in this population remain scarce. Research Design and Methods: We retrospectively assessed the accuracy of the Dexcom G6 CGM system in adults with PBH after Roux-en-Y gastric bypass (RYGB) surgery (n = 70). Glucose excursions were induced using a standardized solid mixed meal test, and reference blood glucose (BG) values were obtained through repeated venous whole blood sampling. CGM accuracy was analyzed separately during stable and dynamic post-meal glucose periods, with dynamic phases stratified according to magnitude and direction of rate of change (RoC). We further estimated the lag time for each sensor and examined predictive factors affecting CGM accuracy. Results: A total of 1822 CGM-BG pairs obtained in 70 subjects were included in the analysis. Mean absolute relative difference (MARD) at stable and dynamic levels were 9.6% and 16.4%, respectively. Post-meal, 67.6% of pairs were within 15% / 15 mg/dL, 78.0% within 20% / 20 mg/dL and 90.8% within 30% / 30 mg/dL. Performance was worse at rapid plasma glucose decline (> 1.5 mg/dL/min) and CGM values at the time of plasma glucose nadir were systematically higher (bias 8.2 mg/dL). Plasma-interstitium time delay was estimated at 9.8 minutes. No subject or sensor characteristic had a significant impact on CGM accuracy. Conclusions: Meal-induced glucose dynamics, particularly rapid declines, challenge CGM accuracy in people with PBH and must be carefully considered when diagnosing or managing the condition.

Fields of Study & Subjects

Bariatric Surgery and Outcomes Diabetes Management and Research Hyperglycemia and glycemic control in critically ill and hospitalized patients

Publication details

Publication date
31 March 2026
Language
EN