Comparison of visibility and material properties of the iCover and VBX stent grafts when used as bridge stents in fenestrated endografts. An in-vitro study
Abstract
BACKGROUND: To in-vitro compare different material properties of two balloon-expandable covered-stents, VBX (W.L. Gore & Associates, Flagstaff, AZ, USA) and iCover (iVascular, Barcelona, Spain), as bridging stents in fenestrated aortic endoprosthesis. METHODS: Four experiments were designed to evaluate a total of 12 stents (6 VBX 6x59 mm and 6 iCover 6x57 mm). Visibility of the stents was evaluated by computed tomography (CT) comparing Hounsfield units (HU) of the radiopaque marks of the balloon, the first and second stent crown and the whole crimped and expanded stent. Material assessment was analyzed performing three experiments: 1) evaluation of the diameter in three different zones, the percentage of stent shortening and its integrity after the expansion and flaring inside a 6-mm-diameter handmade fenestration; 2) evaluation of the residual lumen area of the stent with a 90° angulation; 3) water permeability test; measuring the water entry pressure and the integral water leakage at 120 mmHg. All models underwent a CT scan immersed in a 37 °C water bath and analyzed in the workstation Osirix Imaging Software v 3.8.1, 32 bit (Pixmeo, Geneva, Switzerland) by two different observers. RESULTS: In the crimped configuration, both stents showed statistically significant differences in radiopacity. iCover was more radiopaque than VBX for the entire stent as well as for the first and second stent crowns (4935.23 vs. 2574.39 HU; 723.81 vs. 262.63 HU; and 535.94 vs. 259.02 HU, respectively; P
Fields of Study & Subjects
Coronary Interventions and Diagnostics
Aortic aneurysm repair treatments
Infectious Aortic and Vascular Conditions
Publication details
- Journal / source
- The Journal of Cardiovascular Surgery
- Publisher
- Edizioni Minerva Medica
- Volume
- 67
- Issue
- 1
- Pages
- 107–115
- Publication date
- 25 February 2026
- Language
- EN