Feasibility Study of 18 F-FDG PET Short-Duration Scan in Underweight Patients with Lung Cancer

View at publisher
Authors: Hiroaki Sagara, Go Akamatsu, Chikara Mano, Anri Inaki, Kazumasa Inoue
Cited by: 1
References: 21

Abstract

This retrospective study investigated the feasibility of short-duration 18F-FDG PET/CT imaging for underweight patients with lung cancer. Methods: Short-duration imaging conditions were established so that the noise-equivalent count density satisfied the reference level proposed by the Japanese Society of Nuclear Medicine. Results: Scan time could be reduced from 120 to 52 s/bed position (a 57% reduction) by optimizing Bayesian penalized likelihood (Q.Clear) reconstruction parameters using a β of 650 while preserving image noise levels similar to those in standard-duration imaging (β = 350). Although SUVmax decreased under the shorter-scan condition, the change in SUVpeak was approximately −5% for lesions with diameters of more than 20 mm. Conclusion: The acquisition time for 18F-FDG PET for underweight lung cancer patients can be shortened by 57% without compromising image quality. SUVpeak appears to be a reliable, robust quantitative metric for evaluating larger lesions under such conditions.

Fields of Study & Subjects

Medical Imaging Techniques and Applications Lung Cancer Diagnosis and Treatment Radiomics and Machine Learning in Medical Imaging

Publication details

Journal / source
Journal of Nuclear Medicine Technology
Publisher
Society of Nuclear Medicine
Volume
54
Issue
3
Pages
318–323
Publication date
10 February 2026