Hydrodynamic Characteristics of Bovine Pericardial and Porcine Valves Using a Mock Circulatory System Mimicking the Aortic and Pulmonary Positions
Abstract
Purpose: Aortic prostheses are used in pulmonary positions due to structural similarities between the pulmonary and aortic valves.However, there are no available studies that have comprehensively evaluated the mechanism of bioprosthetic aortic valves under pulmonary conditions.Materials and Methods: Using a mock circulatory system, we evaluated the hydrodynamic characteristics of bovine pericardial and porcine valves.Geometric orifice area, regurgitant and leakage volume, regurgitant fraction, peak pressure gradient, and forward flow volume were evaluated in different pulmonary pressure conditions (from 15/5 mm Hg to 75/35 mm Hg) and normal aortic pressure (110/80 mm Hg).Results: Bovine pericardial valves were associated with larger opening area (0.93±0.01 vs.1.70±0.01 for 23-mm valve; 0.99±0.01vs.1.75±0.01 for 25-mm valve; 1.58±0.01 vs. 2.25±0.02for 27-mm valve; all p