Evaluation of the accuracy of cone-beam CT–based dose calculation for target volumes and organs at risk in left-sided breast cancer radiotherapy

Read full text
Authors: Xiaoxiao Hou, Zhenglu Bai, Xiaoxiao Jin, Erxun Dai, Yaqi Jiang, Jun Ma, J. Li
Cited by: 0
References: 21

Abstract

Introduction Accurate dose calculation is essential in breast cancer radiotherapy. This study aimed to evaluate the feasibility and accuracy of cone-beam CT (CBCT) images with HU–relative electron density (HU-RED) correction for dose calculation in left-sided breast cancer radiotherapy. Methods Twenty postoperative patients receiving adjuvant radiotherapy at Subei People’s Hospital of Jiangsu Province (Yangzhou, China) from January 2022 to December 2024 were retrospectively enrolled. A patient-specific HU-RED calibration curve was generated using an improved density override method. Treatment plans created on planning CT (pCT) were transferred to CBCT for dose recalculation with identical optimization parameters. Dosimetric parameters of the planning target volume (PTV) and organs at risk (OARs) were compared. Two-dimensional gamma analysis was performed to assess dose consistency. Equivalence between CBCT- and pCT-based dose calculations was further evaluated using the two one-sided tests (TOST), Lin’s concordance correlation coefficient (CCC), intraclass correlation coefficient (ICC), and Bland–Altman analysis. Results Differences in PTV dosimetric indices (D2, D50, D98, Dmean, HI, CI) between CBCT- and pCT-based plans were small, with mean deviations 0.05). OAR parameters, including cardiac Dmean, V20, V30, V50 and lung Dmean, V20, V30, also showed minimal variation (

Fields of Study & Subjects

Advanced Radiotherapy Techniques Radiation Dose and Imaging Radiation Therapy and Dosimetry

Publication details

Journal / source
Frontiers in Oncology
Volume
16
Pages
1768319–1768319
Publication date
3 March 2026
Language
EN