Biomechanical Analysis of Zygomatic Implants with Clinically Relevant Configurations: A Finite Element Study

Read full text
Authors: Laura H. J. Alberto, Jason Griggs, Michael D. Roach, Randal S Williamson, Ravi Chandran, Matthew W. Priddy, Yuefeng (Jordan) Lu, Patrick F. Bergin, Yuanyuan Duan
Cited by: 1
References: 33

Abstract

Zygomatic implants have emerged as a promising surgical technique to rehabilitate severe atrophic maxilla and maxillectomy. However, this implant design lacks an appropriate standard test protocol. This study aimed to simulate and evaluate zygomatic implants under compressive and lateral loads using finite element analysis, following the current ISO parameters for dental implants, employing different cylindrical holders. These holders simulate the bone configuration of clinical scenarios of the zygoma anatomic-guided approach (ZAGA) and maxillectomy rehabilitation with zygomatic implants. Six three-dimensional models were generated and subjected to a 150 N compressive load at a 40° angle. In the next step, a simultaneous 50 N lateral load was applied. Von Mises stress data and contour plots were collected and analyzed. Von Mises peak stresses increased by more than 100% when a lateral load was applied. Data demonstrated the influence of the setup support and lateral load on the implant components’ peak stress values.

Fields of Study & Subjects

Dental Implant Techniques and Outcomes Dental Radiography and Imaging Periodontal Regeneration and Treatments

Publication details

Journal / source
Applied Sciences
Publisher
MDPI AG
Volume
16
Issue
3
Pages
1248–1248
Publication date
26 January 2026