Polyethylene Stresses in Lumbar Total Joint Replacement Under Elevated Loading: Insights from an Anatomic Finite Element Model
Abstract
The goal of this study was to assess elevated spinal loading conditions and their effect on the polyethylene stresses of a lumbar total joint replacement (L-TJR). A previously validated lumbar spine finite element model was virtually implanted with an L-TJR at L4-L5 and exposed to three elevated loading conditions: (1) 95th-percentile male body weight while bending forward, (2) combined ±7.5 Nm axial torsion and lateral bending, and (3) ASTM F2423 aggressive loading (1850 N plus 10-12 Nm bending). Combined torsion and lateral bending were considered because these loads and moments may be coupled in demanding real-world scenarios. Across all conditions, contact at the bearing remained confined to the intended spherical surfaces, consistent with Mode I in vitro wear tests, with no evidence of impingement. Contact stresses and von Mises stresses were considered acceptable based on the simulated results of Mode IV impingement tests. Only in one scenario-95th-percentile male body weight with multiaxial torsion-did von Mises stress in the polyethylene slightly exceed the stresses associated with impingement (
Fields of Study & Subjects
Spine and Intervertebral Disc Pathology
Orthopaedic implants and arthroplasty
Cervical and Thoracic Myelopathy
Publication details
- Journal / source
- Bioengineering
- Publisher
- MDPI AG
- Volume
- 13
- Issue
- 1
- Pages
- 66–66
- Publication date
- 7 January 2026
- Language
- EN