Development of a Non-Contact Flow Sensor Based on a Permanent Magnet Metal Clip for Monitoring Circulation Status
Abstract
Foreign matter accumulating on catheters during ascites paracentesis in cancer patients can interfere with the procedure. The paracentesis site must be visually inspected by patients or medical staff. We propose a monitoring method using sensors, as they enable real-time, automatic status detection. The proposed design integrates a sensor into the drainage tube to detect liquid flow using the Lorentz force. The sensor consists of a permanent magnet, a yoke, and a signal processing circuit. Mu-metal shields the magnet to prevent interference with surrounding circuits. Physiological saline solution is used as a substitute for bodily fluids. Sensor performance was evaluated via finite element analysis. The Lorentz force generated an average voltage of 11.07 μV when liquid was present, enabling detection of the flow status. The proposed sensor is non-invasive and features a clip design, allowing attachment and detachment from the drainage tube for reuse. Non-invasiveness ensures safety from infection, and reusability can reduce medical costs. This study proposes a sensor for monitoring peritoneal puncture status. By detecting liquid flow using the Lorentz force, the system enables real-time monitoring during the procedure.
Fields of Study & Subjects
Magnetic Field Sensors Techniques
Non-Invasive Vital Sign Monitoring
Sensor Technology and Measurement Systems
Publication details
- Journal / source
- Biosensors
- Volume
- 16
- Issue
- 2
- Pages
- 78–78
- Publication date
- 27 January 2026
- Language
- EN