Evaluation of Current Injection and Voltage Acquisition Patterns for Electrical Impedance Tomography Image Reconstruction: A Simulation Study

Read full text
Authors: Minh Quan Cao Dinh, Hai Anh Nguyen Thi, Dang Khoa Vo, Lin Dan Lieu, Trung Thach Nguyen, Hong Duyen Trinh Tran
Cited by: 0
References: 10

Abstract

The influence of different voltage measurement and current injection configurations on the quality of image reconstruction in electrical impedance tomography (EIT) was investigated using numerical simulations. Adjacent and opposing techniques were systematically used to examine their effectiveness in voltage acquisition and current delivery. The simulation model employed 16 equally spaced electrodes arranged around a circular domain, with an injected alternating current of 1 mA at a frequency of 50 kHz. A circular object with a conductivity of 0.9 units was sequentially positioned at five distinct locations within the imaging domain, each spaced 0.05 units apart. The reconstructed images were analyzed for positional accuracy and contrast resolution. While each configuration offers specific advantages, they exhibit inherent limitations depending on the application. The results of this study enable the understanding of the trade-offs involved in selecting electrode drive and measurement strategies for optimizing image quality in EIT systems.

Fields of Study & Subjects

Electrical and Bioimpedance Tomography Geophysical and Geoelectrical Methods Microwave Imaging and Scattering Analysis

Publication details

Journal / source
IEEE Eurasia Conference on Biomedical Engineering, Healthcare and Sustainability (ECBIOS)
Pages
20–20
Publication date
2 March 2026