Performance Improvement of the Acrylic Acid–Polyvinyl Alcohol Gel Dosimeter with an Organic Additive for Radiation Oncology Applications

Read full text
Authors: Belal Moftah, Khalid A. Rabaeh, Akram A. Moussa, Abdullah S. Bani Issa, Md A. Al Kafi
Cited by: 1
References: 52

Abstract

This study reports the first preparation and characterization of an acrylic acid–glucose–polyvinyl alcohol (ACAGLPVA) polymer gel dosimeter incorporating glucose as an organic additive for radiation oncology applications. Five formulations with glucose concentrations of 0, 10, 20, 25, and 30 wt% were irradiated using a 6-MV photon beam at doses of 0–60 Gy, and the transverse relaxation rate (R2) was measured by nuclear magnetic resonance (NMR) relaxometry. The optimal formulation (25 wt% glucose) demonstrated an excellent linear dose response between 0 and 30 Gy (R2 = 0.9979) with a sensitivity of 0.177 s−1 Gy−1, followed by a non-linear response at 30–60 Gy. The dosimeter exhibited dose rate independence (200–600 cGy/min), energy independence (6–15 MV), temperature independence (5–35 °C), and post-irradiation stability for at least 7 days. These characteristics demonstrate the potential of ACAGLPVA gel dosimeters for accurate three-dimensional dose verification in modern radiotherapy applications.

Fields of Study & Subjects

Advanced Radiotherapy Techniques Radiation Therapy and Dosimetry Radiation Effects and Dosimetry

Publication details

Journal / source
Gels
Publisher
MDPI AG
Volume
12
Issue
2
Pages
176–176
Publication date
17 February 2026
Language