An Open-Source Horizontal Strabismus Simulator as an Evaluation Platform for Monocular Gaze Estimation Using Deep Learning Models
Abstract
Strabismus affects 2-4% of the global population, with horizontal cases accounting for more than 90%. Automated screening using monocular gaze estimation technology shows promise for early detection. However, existing models assume normal binocular vision, and their applicability to strabismus remains unvalidated due to the lack of evaluation platforms capable of reproducing disconjugate eye movements with known ground-truth angles. To address this gap, we developed an open-source, low-cost (approximately 200 USD) horizontal strabismus simulator. The simulator features two independently controllable artificial eyeballs mounted on a two-axis gimbal mechanism with servo motors and gyro sensors for real-time angle measurement. Mechanical accuracy achieved a mean absolute error of less than 0.1° across all axes, well below the clinical detection threshold of 1 prism diopter (≈0.57°). An evaluation of three representative AI models (Single Eye, GazeNet, and EyeNet) revealed estimation errors of 6.44-8.75°, substantially exceeding the clinical target of 2.8°. At this error level, small-angle strabismus (
Fields of Study & Subjects
Gaze Tracking and Assistive Technology
Ophthalmology and Visual Impairment Studies
Ophthalmology and Eye Disorders
Publication details
- Journal / source
- Journal of Eye Movement Research
- Publisher
- MDPI AG
- Volume
- 19
- Issue
- 1
- Pages
- 20–20
- Publication date
- 9 February 2026