False-Positive IMMY® Serum Cryptococcal Antigen Lateral Flow Assay Results Due to Improper Use of Titration Buffer

Read full text
Authors: Asmus Tukundane, Wilber Bakka, Mercy Amanige, Dora Babirye, Timothy Mugabi, Abdu K Musubire, David B. Meya, David R. Boulware, Brian Doherty, Richard Kwizera, Caleb P Skipper
Cited by: 0
References: 11

Abstract

detection, but false-positive results may arise from procedural or reagent-related errors, underscoring careful operation of the assay to ensure diagnostic accuracy and prevent unnecessary treatment. Two patients who were initially reported as CrAg-positive by a peripheral laboratory were referred to Kiruddu Hospital in Kampala, Uganda, for clinical assessment and confirmatory testing. Repeat tests were conducted using specimen diluent following the manufacturer's protocol, resulting in negative results. Semi-quantitative CrAg LFA testing and a series of control assays were performed to identify the source of error. We were able to consistently reproduce positive results when the titration diluent was inappropriately used instead of the specimen diluent. Serial dilutions confirmed persistent false positivity up to 1:80 when inappropriately using the titration diluent, while all dilutions that appropriately started with the specimen diluent remained negative. We hypothesize that the incorrect use of titration diluent instead of specimen diluent led to false-positive CrAg LFA results due to the absence of the blocking agent that neutralizes heterophilic antibodies. Procedural errors can lead to diagnostic misinterpretation and serious consequences in clinical management, emphasizing the importance of adherence to manufacturer's instructions.

Fields of Study & Subjects

Fungal Infections and Studies Clinical Laboratory Practices and Quality Control Antifungal resistance and susceptibility

Publication details

Journal / source
Journal of Fungi
Volume
12
Issue
2
Pages
135–135
Publication date
12 February 2026
Language
EN