Non-Dispersive Gas Analyzer for H2O and CO2 Flux Analysis by the Eddy Covariance Method

Read full text
Authors: И. Л. Фуфурин, Ivan Karpov, А. П. Костерова, Viacheslav Bessonov, Alexis Yaroslavtsev, Ivan Seregin, Andrey Morozov
Cited by: 0
References: 16

Abstract

This article presents the stages of development of a non-dispersive infrared (NDIR) open-type gas analyzer prototype (NGAP-1) with a fundamentally new discrete IR radiation generation scheme using pulse-width modulation for measuring the dynamics of water vapor and carbon dioxide concentrations, with further application in the instrument base of an ecological and climatic station to implement the eddy covariance (EC) method. In addition, selecting the component base for NGAP-1, as well as its calibration and experimental field validation as part of an ecological and climatic station, are described.

Fields of Study & Subjects

Spectroscopy and Laser Applications Advanced Measurement and Detection Methods Plant Water Relations and Carbon Dynamics

Publication details

Journal / source
Sensors
Volume
26
Issue
2
Pages
560–560
Publication date
14 January 2026