Design and Validation of a Pressure-Driven Liquid Metering System with Heated PTFE Tubing for Laboratory Automation

Read full text
Authors: Joonki Baek, Taegyun Kim, Seungwon Jeong, Ikhyun Kim, Shin Hum Cho, Sungkeun Yoo
Cited by: 1
References: 32

Abstract

This paper presents a pressure-driven liquid transfer system for laboratory automation, along with a physics-based model and calibration method. The device maintains near-isothermal transport by storing reagents at a prescribed temperature and routing the flow through a single PTFE tube enclosed within a heated jacket. The pressure-drop model accounts for temperature-dependent viscosity and the thermal expansion of PTFE. Residual deviations from the no-slip prediction in submillimeter tubing are represented by an effective slip length, which is identified through linear regression. This parameter is subsequently used to calculate the pressure required to achieve a target flow rate. Experimental results compare unheated and heated operating conditions and characterize the dependence of slip length on temperature and flow rate. Under heated operation with slip-compensated pressure commands, the system achieved dispensing accuracy within ±4% over the tested range, whereas unheated operation exhibited larger errors due to axial temperature gradients. These results demonstrate that effective thermal management and slip compensation are critical for accurate pressure-based metering under temperature-sensitive conditions, as validated using water-based tests.

Fields of Study & Subjects

Flow Measurement and Analysis Intravenous Infusion Technology and Safety Water Systems and Optimization

Publication details

Journal / source
Sensors
Publisher
MDPI AG
Volume
26
Issue
2
Pages
700–700
Publication date
21 January 2026
Language
EN