Towards Real-Time Aquatic Monitoring of Strontium-90: Performance Evaluation of CaF2(Eu) and ZnSe(Al,O) Scintillators

Read full text
Authors: Arjana Kolnikaj, Kelum A. A. Gamage, Olaoluwa Popoola, James Graham, Antonio Di Buono
Cited by: 1
References: 25

Abstract

A compact, in situ beta-spectroscopy approach for real-time monitoring of Strontium-90 (Sr-90) in contaminated groundwater has been investigated. Two inorganic scintillators, CaF2(Eu) and ZnSe(Al,O), were coupled to silicon photomultipliers (SiPMs) and evaluated experimentally using custom front-end electronics. This was also modelled with Monte Carlo simulations using the Geant4 toolkit. Although simulations correctly predicted ZnSe(Al,O) has an advantage due to its higher light yield and optical transport, experimental measurements additionally revealed practical limitations of the readout electronics which were not captured in the simulation model. ZnSe(Al,O) showed excellent agreement with the simulated detector response (R2 ≈ 0.86; χ2/NDF ≈ 27). It also attains a higher relative detection efficiency (∼61.5%), yielding faithful capture of the composite Sr-90/Y-90 spectrum with only minor suppression at the extreme high-energy tail. CaF2(Eu) exhibits a deficit at low-mid energies and an apparent enhancement in the high-energy tail. This is consistent with threshold and photon-statistics losses and leads to poorer agreement with simulation (χ2/NDF ≈ 179) and lower overall efficiency (∼22.7%). These findings identify ZnSe(Al,O) as the stronger candidate for an underwater, in situ Sr-90 beta-spectroscopy system and motivate targeted optimisation of SiPM coupling and crystal-edge reflectivity in future designs.

Fields of Study & Subjects

Radiation Detection and Scintillator Technologies Radioactive contamination and transfer Luminescence Properties of Advanced Materials

Publication details

Journal / source
Sensors
Publisher
MDPI AG
Volume
26
Issue
3
Pages
900–900
Publication date
29 January 2026
Language
EN