Measuring Dynamic Tendon Torsion Using Ultrasound Speckle Tracking: Validation with Silicone Phantom and In Vivo Application on Human Tibialis Posterior Tendon

Read full text
Authors: Kun-Lin Hung, De-Kai Syu, Wei-Ning Lee, Pei‐Yu Chen, Chen‐Chie Wang, Wen-Siang Chen, Che-Yu Lin, Hsing‐Kuo Wang
Cited by: 0
References: 72

Abstract

The torsional characteristics of human tendons are recognized to have functional and clinical relevance, but are underexplored due to the limited in vivo assessment methods available to measure the dynamic torsion characteristics of a tendon during movement. This study aimed to validate the use of transverse plane ultrasound speckle tracking (ST) for measuring dynamic torsion on silicone phantoms, and to evaluate the capability and reliability of ST in measuring dynamic torsion of the human tibialis posterior tendon (TPT) in vivo. Of the ten silicone phantoms tested in the validation study, ST measurement results strongly correlated with the referencing marker tracking method (R2 = 0.81–0.95) and had measurement error similar to or smaller than the hypothesized accuracy of 3° (p > 0.045). Subsequently, when ST was applied to nineteen healthy participants’ TPT in vivo, it was capable of characterizing the dynamic external torsion of the TPT during 0–20° passive foot pronation. Strong correlations were found between the ST-measured angle and the foot pronation angle (R2 = 0.98–0.99), and the test–retest reliability was moderate to good (ICC = 0.73–0.87). These findings suggested that ST is a valid and reliable method for measuring dynamic tendon torsion characteristics.

Fields of Study & Subjects

Ultrasound Imaging and Elastography Tendon Structure and Treatment Lower Extremity Biomechanics and Pathologies

Publication details

Journal / source
Sensors
Volume
26
Issue
4
Pages
1187–1187
Publication date
11 February 2026
Language
EN