Single VDCC-Based Mixed-Mode First-Order Universal Filter and Applications in Bio-Signal Processing Systems
Abstract
This paper presents a compact mixed-mode first-order universal filter based on a single voltage differencing current conveyor (VDCC), which can function in all four possible operation modes, i.e., voltage mode (VM), trans-admittance mode (TAM), current mode (CM), and trans-impedance mode (TIM). The proposed configuration requires only two grounded resistors and one floating capacitor, which contributes to a low component count, facilitates integration, and allows for the electronic tunability of the pole frequency through the transconductance gain of the VDCC. This work also demonstrates two practical biomedical applications: an electrocardiogram (ECG) acquisition system utilizing the VM low-pass filter for noise suppression and a bioimpedance (BioZ) measurement system employing the proposed configuration-based CM oscillator circuit as a sinusoidal excitation source. The performance validation confirms the accuracy of impedance extraction and the preservation of waveforms using tissue-equivalent models. The results demonstrate that the proposed VDCC-based filter offers a compact, power-efficient, and versatile analog signal-processing solution suitable for modern biomedical instrumentation.
Fields of Study & Subjects
Analog and Mixed-Signal Circuit Design
ECG Monitoring and Analysis
Electrical and Bioimpedance Tomography
Publication details
- Journal / source
- Technologies
- Publisher
- MDPI AG
- Volume
- 14
- Issue
- 2
- Pages
- 101–101
- Publication date
- 4 February 2026
- Language
- EN