Stent geometry achieved by stepwise provisional stenting in left main trifurcation: Bench-test comparison of different side-branch intervention techniques

Read full text
Authors: A. Mâaroufi, Mattia Lunardi, Judit Andréka, Enrico Romagnoli, Johannes Gollmer, Cristina Aurigemma, Lazzaro Paraggio, Francesco Bianchini, G P Plank, Anton J. Prassl, Carlo Trani, G G Toth
+1 more Francesco Burzotta
Cited by: 0

Abstract

BACKGROUND: Left main coronary trifurcations pose specific challenges for percutaneous coronary intervention due to complex anatomy and lack of standardized approaches. AIMS: We assess to evaluate the stent conformations achieved by different adaptations to trifurcations of stepwise provisional stenting in a bench test. METHODS: Percutaneous coronary intervention procedures were performed in silicon trifurcation models using standard equipment. After main vessel stenting and proximal optimization technique, side branch (SB) rewiring was conducted via same-cell (SAC) or neighboring-cell (NEC) under optical coherence tomography guidance. SBs dilation was then performed using either simultaneous trissing or serial kissing balloon techniques. Following repeat proximal optimization technique, stent expansion and apposition were assessed with optical coherence tomography and micro-computed tomography. RESULTS: All 4 technique combinations were feasible and resulted in satisfactory stent expansion and apposition. Micro-computed tomography analysis showed that trissing improved expansion at the polygon of confluence compared to serial kissing (mean 23.7 vs. 21.1 mm²; P = 0.043), while SAC rewiring yielded better expansion at the distal main vessel ostium than NEC (mean 10.3 vs. 9.6 mm²; P = 0.003). Rewiring technique determined the conformation of the stent at the level of SBs takeoff: NEC produced 2 stent fenestrations and SAC a single stent fenestration (with a minimal area significantly smaller than the sum of the 2 achieved by NEC: 16.4 mm² vs. 27.9 mm²; P = 0.019). CONCLUSIONS: All provisional stenting strategies tested were technically feasible for left main trifurcations. SB rewiring and ballooning technique influenced stent configuration, particularly the number and size of fenestrations, but had limited impact on overall stent expansion and apposition.

Fields of Study & Subjects

Coronary Interventions and Diagnostics Cardiac Imaging and Diagnostics Acute Myocardial Infarction Research

Publication details

Journal / source
Kardiologia Polska
Volume
84
Issue
3
Pages
356–364
Publication date
16 February 2026
Language
EN