Proposed S‑Parameter Method to Measure the Voltage Division Factor of a LISN

Read full text
Authors: Michael Heckrotte, Zhong Chen
Cited by: 0

Abstract

The voltage division factor (VDF) of a line impedance stabilization network (LISN) is also known as LISN insertion loss (LISNIL). VDF is defined differently than the usual RF insertion loss parameter; hence, VDF is measured using a unique setup with a three‑port coaxial T‑connector. We describe and propose an S‑parameter-based method of measuring VDF. Using this method, VDF can be measured simultaneously with LISN input impedance by acquiring both forward S‑parameters and performing a simple calculation. We derive model equations based on ideal instrumentation and show the equivalence of the two methods. Comparative measurements performed on representative LISNs experimentally confirm the mathematical equivalence. Thus, we show that the T‑connector method is redundant during LISN calibrations.

Fields of Study & Subjects

Electromagnetic Compatibility and Noise Suppression Electromagnetic Compatibility and Measurements Advanced Electrical Measurement Techniques

Publication details

Publication date
13 January 2026
Document type
Preprint