Tailored Calibration Approach for Hemodynamic Boundary Conditions in CT-FFR
Abstract
BACKGROUND AND OBJECTIVES: Computed tomography-derived fractional flow reserve (CT-FFR) uses standardized boundary conditions that cannot represent individual patient characteristics. We developed a calibration method to individualize boundary conditions using invasive fractional flow reserve (FFR) measurement. This study assessed the feasibility and impact on prediction of lesion-specific pressure drops (ΔFFR), post-percutaneous coronary intervention (PCI) FFR, and high-risk plaque (HRP) discrimination. METHODS: This retrospective multicenter study analyzed 79 patients with single left anterior descending lesions. Lumped parameter models were used to calibrate optimal boundary conditions by iteratively minimizing differences between invasive FFR and CT-FFR. HRP was assessed using an artificial intelligence-enabled analysis system, defined as low attenuation plaque, positive remodeling, or both. Discrimination performance was evaluated using ΔCT-FFR and CT-FFR gradient. RESULTS: In 79 patients (median age 65.5 years, 77.2% male), median angiographic percent diameter stenosis 60.0%, median invasive FFR was 0.83 and CT-FFR was 0.81 before calibration. Following calibration, FFR correlation improved from 0.590 to 0.997 (p
Fields of Study & Subjects
Cardiac Imaging and Diagnostics
Advanced X-ray and CT Imaging
Cardiovascular Function and Risk Factors
Publication details
- Journal / source
- Korean Circulation Journal
- Publisher
- XMLink
- Volume
- 57
- Publication date
- 5 August 2026
- Language
- EN