CIRCLE software for creation of corneal side cut in small-incision lenticule extraction

Read full text
Authors: Gitansha Shreyas Sachdev, Shreyas Ramamurthy, Ujjwala Pandey
Cited by: 0
References: 10

Abstract

We report an innovative use of the CIRCLE software (Carl Zeiss, Meditec) to manage a case of suction loss during small-incision lenticule extraction (SMILE). We describe a case of unstable suction during the corneal side-cut incision, wherein an immediate re-hold was obtained on the cornea, and the femtosecond delivery was completed. However, the side-cut incision was formed in the paracentral anterior cornea. The CIRCLE software (Pattern D) was used to fashion a new corneal side cut and gain subsequent access to the lenticule. The patient achieved an uncorrected distance visual acuity of 20/20 on postoperative day one. We present a rare case where the machine was able to re-establish suction after an excessive eye movement, while the laser delivery proceeded to continue uninterrupted. CIRCLE software provided a unique and novel application for gaining access to the SMILE lenticule in the case of failure of corneal side-cut formation in the intended location.

Fields of Study & Subjects

Corneal surgery and disorders Intraocular Surgery and Lenses Ocular Infections and Treatments

Publication details

Journal / source
Indian Journal of Ophthalmology - Case Reports
Volume
6
Issue
1
Pages
228–230
Publication date
1 January 2026