A 2.5 $μ$W 30 nV/$\surd$Hz Instrumentation Amplifier for Bioimpedance Sensors with Source Degenerated Current Mirror and DTMOS Transistor

Read full text
Authors: Yu Xue, Kwantae Kim
Cited by: 0

Abstract

This paper proposes a low-power and low-noise instrumentation amplifier (IA) tailored for bioimpedance sensing applications. The design originates from a gain-boosted flipped voltage follower (FVF) transconductance (TC) stage and integrates two complementary circuit techniques to improve the noise performance. To achieve an optimal balance between input-referred noise and available voltage headroom, a source-degenerated current mirror (SDCM) is adopted, resulting in reducing the input-referred noise by 7.95% compared with a conventional current mirror structure. In addition, a dynamic threshold MOSFET (DTMOS) scheme is employed to enhance the effective transconductance, leading to a further 11.66% reduction in input-referred noise. Simulated in a 28 nm CMOS process demonstrate that the proposed IA achieves an input-referred noise floor of 30 nV/$\surd$Hz and a bandwidth of 1.44 MHz, while consuming only 2.5 $μ$W from a 0.8 V supply. Compared to the baseline design, the proposed approach achieves a 32.4% reduction in power consumption without degrading noise performance. The complete design parameters are open-sourced in this paper, to ensure reproducibility and facilitate future developments.

Fields of Study & Subjects

Analog and Mixed-Signal Circuit Design Electrical and Bioimpedance Tomography Innovative Energy Harvesting Technologies

Publication details

Journal / source
arXiv (Cornell University)
Publication date
3 January 2026
Document type
Preprint