Clinical assessment and acceptance criteria for continuous glucose monitoring (CGM) system performance: A proposed guideline by the IFCC Working Group on CGM.
Abstract
Continuous glucose monitoring (CGM) systems are essential tools for modern diabetes care. However, the assessment of their analytical performance has been hindered by a lack of standardization. This poses challenges for the field, as the choice of study procedures and evaluation methods can influence the observed performance, further complicating the comparisons between CGM systems and confounding the interpretation and use of CGM metrics as research outcomes. To address this issue, the Working Group on CGM of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) has developed a comprehensive guideline for the clinical assessment of CGM systems. The guideline defines requirements for study design and procedures, characteristics of comparator measurements, minimum requirements for accuracy, and performance characterization.
Fields of Study & Subjects
Diabetes Management and Research
Hyperglycemia and glycemic control in critically ill and hospitalized patients
Spectroscopy Techniques in Biomedical and Chemical Research
Publication details
- Journal / source
- Open Access CRIS of the University of Bern
- Publication date
- 30 January 2026
- Language
- EN