Data from: Refillable intraocular drug-eluting implant

Read full text
Authors: Charles M. T. DeBoer, Hyeonji Kim, Nae-Won Kang, Roger Wise, Akira Feliz, David J. Myung
Cited by: 0

Abstract

The increasing incidence of ophthalmic diseases has led to the need for multiple dosage regimens, which can cause patient discomfort and noncompliance due to the frequency of medication administration. Still, ophthalmic drug delivery methods, such as eye drops and intravitreal injection, pose challenges, such as poor bioavailability, short half-life, and patient compliance. In this study, we introduce a novel refillable intracapsular drug-eluting reservoir for the long-term management of ocular diseases. This device, made of medical-grade silicone and stainless steel, has features to contour the lens capsule to facilitate microincisional implantation. Rheological and mechanical analyses of the silicone revealed the optimized conditions for the device construction and its application compliance for ocular intracapsular implantation. Furthermore, in vitro release studies exhibit controlled drug release, the kinetics of which were consistent with Fickian diffusion, while ex vivo implantation testing shows that the device can be easily delivered and placed at the time of cataract surgery. Taken altogether, the developed implant holds significant potential for improving therapeutic outcomes while offering a practical surgical application and compliance of patients.

Fields of Study & Subjects

Advanced Drug Delivery Systems Intraocular Surgery and Lenses Hydrogels: synthesis properties applications

Publication details

Journal / source
DRYAD
Publication date
8 February 2026
Language
EN
Document type
Dataset