Rethinking soil incubation greenhouse gas monitoring: How the Picarro G2508 coupled with the Sage gas autosampler compares against gas chromatography analysis
Abstract
High‑precision greenhouse gas (GHG) measurements are essential for accurately assessing soil carbon and nitrogen cycling. Traditionally, gas chromatography (GC) with autosamplers has been the standard for soil incubation studies due to its accuracy and high‑throughput capability. Recent advances in laser‑based systems, such as Picarro’s G2508 Cavity Ring‑Down Spectroscopy (CRDS) analyzer, now enable continuous, real‑time monitoring of CO₂, CH₄, N₂O, and other gases. When paired with Picarro’s new Sage gas autosampler, the system supports automated, high‑throughput measurements of discrete, small‑volume gas samples using only zero air (or N₂) for flushing and requiring minimal maintenance.To evaluate performance relative to GC, we conducted an intercomparison study using 60 mL samples split into two equal aliquots—one measured with the G2508–Sage system and one with GC. Certified reference gases (9.9 ppm N₂O, 10 ppm CH₄, 1008 ppm CO₂) and their 10–80% dilutions were analyzed. Both systems achieved coefficients of variation (CVs) below 5%, with the G2508–Sage consistently showing lower CVs when sample concentrations were within the analyzer’s dynamic range. Strong linear correlations (R² > 0.99) were observed across all gases.For soil incubation headspace samples containing elevated GHG concentrations, the two systems generally agreed within CVs
Fields of Study & Subjects
Spectroscopy and Laser Applications
Soil Carbon and Nitrogen Dynamics
Atmospheric and Environmental Gas Dynamics
Publication details
- Journal / source
- EGU General Assembly Conference Abstracts
- Publication date
- 13 March 2026
- Language
- EN