Synthetic Slanted-Edge Image Generator for MTF Validation
Abstract
This script generates a synthetic slanted-edge image for modulation transfer function (MTF) validation in radiographic imaging systems. A binary edge is analytically defined on a highly oversampled grid and then binned to detector resolution, providing accurate pixel-aperture integration without large memory usage. An optional Gaussian blur can be applied in physical units to model additional system blur. The output is a 32-bit floating-point TIFF image suitable for use as a reference input in standard slanted-edge MTF analysis pipelines.
Fields of Study & Subjects
Digital Radiography and Breast Imaging
Image and Video Quality Assessment
Radiation Dose and Imaging
Publication details
- Journal / source
- Zenodo (CERN European Organization for Nuclear Research)
- Publication date
- 12 January 2026
- Language
- EN