Development and Validation of an RP-UHPLC Method for Quantitative Determination of Eugenol as per ICH Guidelines

Read full text
Authors: Dr. Jitendra Banweer Pooja Tilak* Manish Sharma
Cited by: 0

Abstract

In the present research work, a successful attempt was made for “Analytical method development and validation for the determination of eugenol by RP- UHPLC method.The RP-UHPLC-system with Hypersil Gold C18 column (150mm×4.6mm, particle size 3μ) was used in this study. Acetonitrile and water in 60:40 (v/v) ratio was chosen as the isocretic mobile phase under a column temperature of 25ºC. The detection wavelength was set at 285 nm with a flow rate of 0.8 mL/min. The retention times of eugenol was obeserved 4.93 min. which shows the developed method to be cost effective, Rapid, Simple and Accurate and can be successfully employed for the the concurrent estimation of eugenol.Praposed RP-UHPLC method was found to be linear in the range of 2-10 μg/ml. and the correlation coefficient found to be 0.999. The method showed good intraday precision (RSD 0.269) and interday precision (RSD 0.411). The DL and QL were found to be 0.5 and 0.2 μg mL−1, respectively. Method validation was performed according to the International Conference on Harmonization guidelines

Fields of Study & Subjects

Analytical Methods in Pharmaceuticals Analytical Chemistry and Chromatography Chromatography in Natural Products

Publication details

Journal / source
Zenodo (CERN European Organization for Nuclear Research)
Publication date
26 February 2026
Language
EN