High-resolution melting (HRM) analysis of the three designed primer pairs was performed using the BIOER LineGene MiniS tool (Hangzhou, China). The derived HRM curves showed that fluorescence was most frequently detected between 56.75°C and 58.7°C, indicating that the optimal annealing temperature for primers N1, N2, and N3 is within this range. The presence of a single melting curve suggests that the primers specifically amplify the target sequence. Based on Ct values, although the differences were not significant, primer N3 showed slightly better amplification efficiency due to its lower Ct value.