Abstract
BACKGROUND: We evaluated pressures and tightening-system use of 3.8cm-wide, self-securing-strap/redirect-and-tightening-system X8T-T2G tourniquets (X8Ts). METHODS: Applied left/right, mid-arm and mid-thigh, 40 recipients, first completion one-click-past arterial occlusion, release 100s later. Compared to concurrent study: four 3.8cm-wide, windlass-rod-tightening-system tourniquets (n=30 each). RESULTS: All X8T reached occlusion and completion without securing struggles. Maintaining occlusion till release, 3 arms/25 thighs required additional click once, 2/2 required twice, and 1 thigh required thrice. Nineteen arms large enough for X8T pressure monitoring: occlusion median 298mmHg (minimum 188, interquartile range 262, 322, maximum 335), first completion 338mmHg (219, 310, 372, 391), pre-release 304mmHg (172, 270, 323, 342). Thigh pressures: occlusion 359mmHg (273, 334, 388, 478), first completion 393mmHg (308, 366, 426, 489), pre-release 368mmHg (289, 339, 390, 462); each< respective windlass-rod tourniquet pressures (every P