Use of Diluted Rapid-Acting Insulin in the Minimed™ 780G Automated Insulin Delivery System in a Young Child with Type 1 Diabetes
Abstract
The MiniMed™ 780G automated insulin delivery (AID) system improves glycemic control in type 1 diabetes (T1D), but Auto Mode requires a total daily insulin dose (TDI) ≥8 U, limiting its use in young children or individuals in the honeymoon phase. We report a 2.5-year case of a child with T1D and TDI < 8 U successfully treated with diluted lispro insulin (1:1 with 0.9% NaCl) delivered via continuous subcutaneous insulin infusion in the 780G AID system. Auto Mode functioned stably, and the boy maintained excellent metabolic control: Time in Range > 80%, Time in Tight Range > 60%, Coefficient of Variation < 33%, and Glucose Management Indicator < 6.5%, with no clinically significant hypoglycemia. This case demonstrates that insulin dilution can safely enable AID in people with diabetes with very low TDI.
Fields of Study & Subjects
Diabetes Management and Research
Diabetes and associated disorders
Diabetes Management and Education
Publication details
- Journal / source
- Clinical Diabetology
- Volume
- 15
- Issue
- 2
- Pages
- 85–87
- Publication date
- 6 March 2026
- Language
- EN