Kahook Dual-Blade goniotomy combined with phacoemulsification in primary open-angle glaucoma patients: long-term results

Read full text
Authors: Antônio Lucas Oliveira Correia, Breno Santos de Holanda, Vital Paulino Costa, José Paulo Cabral de Vasconcellos
Cited by: 0

Abstract

PURPOSE: Glaucoma is a chronic and progressive disease that requires long-term treatment and continuous monitoring. The Kahook Dual Blade, a device used to perform goniotomy in adults, is designed to improve intraocular pressure control in patients with glaucoma. This study aimed to evaluate the long-term efficacy and safety of kahook dual blade goniotomy in glaucoma patients undergoing cataract surgery over a 36-month follow-up. METHODS: This was a retrospective case series including 56 eyes from 56 patients with mild-to-moderate primary open-angle glaucoma who underwent phacoemulsification combined with kahook dual blade goniotomy. Mean intraocular pressure values, number of preoperative and postoperative hypotensive eye drops, procedure survival, and complications were evaluated over 36 months. Surgical success was defined as either a reduction in intraocular pressure of ≥20% with intraocular pressure between 6 and 18 mmHg without additional medication or a reduction of ≥1 eye drop with intraocular pressure between 6 and 18 mmHg. RESULTS: The mean preoperative intraocular pressure decreased from 15.96 ± 2,83) mmHg to 13.14 ± 2,11) mmHg at 36 months, representing a 14.9% reduction (p

Fields of Study & Subjects

Glaucoma and retinal disorders Intraocular Surgery and Lenses Corneal surgery and disorders

Publication details

Journal / source
Arquivos Brasileiros de Oftalmologia
Volume
89
Issue
1
Pages
1–6
Publication date
1 January 2026
Language
EN