Analytical transfer characteristics of axially symmetric 3D optical imaging systems
Abstract
This paper presents a unified analytical framework to derive the three-dimensional transfer characteristics of axially symmetric optical imaging systems operating in the far field, with emphasis on coherence scanning interferometry, confocal microscopy, and focus variation techniques. While many 3D optical instruments are treated as linear systems, in practice inconsistencies remain in how their transfer functions are derived and interpreted, particularly across forward- and backscatter geometries. Addressing this, we develop closed-form expressions for the 3D transfer function and point spread function for generic back and forward-scatter systems under commonly applied apodization conditions (e.g., uniform, root-cosine and cosine). These derivations clarify the spatial frequency support and resolution trade-offs intrinsic to each geometry and validate the characteristic “bowtie” and “umbrella” structures observed experimentally in the spatial frequency domain. Our results not only resolve ambiguities in previous numerical models but also provide a rigorous analytical foundation for performance assessment and instrument design. The formalism is robust, generalizable, and directly applicable to the modeling and correction of real instruments in 3D surface metrology and optical tomography.
Fields of Study & Subjects
Optical measurement and interference techniques
Surface Roughness and Optical Measurements
Digital Holography and Microscopy
Publication details
- Journal / source
- Figshare
- Publication date
- 1 January 2026