Cryoablation temperature monitoring with dense ultrasonic speed-of-sound shift imaging

View at publisher
Authors: Gaya Lamm, Tal Grutman, Mike Bismuth, Tali Ilovitsh
Cited by: 0
References: 68

Abstract

Abstract Accurate temperature monitoring during cryoablation, a minimally invasive technique that destroys tissue locally by forming an ice ball around an inserted cryoprobe, is vital for achieving complete ablation while protecting surrounding tissue. We present a dense slowness-shift imaging method that estimates local speed-of-sound changes from ultrasound B-mode images using optical flow. This single-transducer, image-based approach enables mapping of spatial temperature change without requiring additional hardware. Cryoablation experiments in a tissue-mimicking phantom and ex vivo turkey breast demonstrated that slowness deviation increases with decreasing temperature. In the phantom, the dependence was linear (α a = -20.70 ηs·m −1 C° −1 ), while in turkey breast it presented an exponential relationship (α t = 34.04×exp(0.075(-ΔT)) ηs·m −1 C° −1 ). The algorithm detected sub-degree temperature variations and accurately tracked cooling down to -39.4 ± 5.6 °C. This work demonstrates the feasibility of ultrasound-based, noninvasive temperature monitoring during cryoablation, providing a scalable, real-time alternative to existing invasive or high-cost thermal assessment techniques.

Fields of Study & Subjects

Ultrasound Imaging and Elastography Photoacoustic and Ultrasonic Imaging Ultrasound and Hyperthermia Applications

Publication details

Journal / source
bioRxiv (Cold Spring Harbor Laboratory)
Publication date
4 January 2026
Document type
Preprint