Cryoablation temperature monitoring with dense ultrasonic speed-of-sound shift imaging
Abstract
Abstract Accurate temperature monitoring during cryoablation, a minimally invasive technique that destroys tissue locally by forming an ice ball around an inserted cryoprobe, is vital for achieving complete ablation while protecting surrounding tissue. We present a dense slowness-shift imaging method that estimates local speed-of-sound changes from ultrasound B-mode images using optical flow. This single-transducer, image-based approach enables mapping of spatial temperature change without requiring additional hardware. Cryoablation experiments in a tissue-mimicking phantom and ex vivo turkey breast demonstrated that slowness deviation increases with decreasing temperature. In the phantom, the dependence was linear (α a = -20.70 ηs·m −1 C° −1 ), while in turkey breast it presented an exponential relationship (α t = 34.04×exp(0.075(-ΔT)) ηs·m −1 C° −1 ). The algorithm detected sub-degree temperature variations and accurately tracked cooling down to -39.4 ± 5.6 °C. This work demonstrates the feasibility of ultrasound-based, noninvasive temperature monitoring during cryoablation, providing a scalable, real-time alternative to existing invasive or high-cost thermal assessment techniques.
Fields of Study & Subjects
Ultrasound Imaging and Elastography
Photoacoustic and Ultrasonic Imaging
Ultrasound and Hyperthermia Applications
Publication details
- Journal / source
- bioRxiv (Cold Spring Harbor Laboratory)
- Publication date
- 4 January 2026
- Document type
- Preprint