Abstract Introduction The 2024 McDonald criteria incorporate the central vein sign (CVS) and paramagnetic rim lesions (PRL) as supportive imaging biomarkers for MS diagnosis. While susceptibility-weighted-imaging (SWI) and T2*-weighted echo-planar-imaging (EPI) are generally used to assess CVS/PRL, their relative performance remains unclear. This study compared high-resolution isotropic-T2*-EPI with non-isotropic SWI for CVS/PRL detection. Materials and Methods In this multi-centre study, 21 patients with MS underwent harmonized 3T-MRI including EPI and SWI. CVS and PRL were evaluated according to NAIMS criteria. Whole-brain and controlled lesion analyses on 120 pre-selected lesions were performed independently for each contrast, with EPI serving as reference standard. Results In whole-brain analyses, SWI showed good sensitivity for CVS eligibility and positivity (AC1=0.68-0.78) but significant directional disagreement with EPI (p