An LC-MS/MS method development and validation for the quantification of selpercatinib in spiked human K2EDTA plasma

Read full text
Authors: Prathusha Katthala, G. Shiva Kumar
Cited by: 0

Abstract

An LC-MS/MS method is needed to accurately measure the concentration of selpercatinib in a biological sample.This method should be both specific and sensitive.The processed materials were separated using a Zorbax SB-C18 column (50 mm × 4.6 mm) 3.5 µm with a mobile phase of acetonitrile, methanol, and 0.1% HCOOH in the proportion of 20:70:10.The mobile phase was measured using a column with 0.7ml/minute rate of flow.The drug and internal standard, ibrutinib were assessed by monitoring the transitions of m/z -526.25/450.16 and 441.20/55.01 for selpercatinib and IS, correspondingly, in multiple reaction monitoring mode.The linearity line and correlation coefficient (r2) values were y = 0.0012x -0.0378 and >0.99, correspondingly.For the QC samples (253, 1,800, and 2,700 ng/ml), the %RSD of inter and intraday precision of the developed method ranged from 2.41% to 5.23%.The proposed method was developed and validated as per the regulatory guidelines.

Fields of Study & Subjects

Analytical Methods in Pharmaceuticals Biosimilars and Bioanalytical Methods Forensic Toxicology and Drug Analysis

Publication details

Journal / source
Journal of Applied Pharmaceutical Science
Volume
16
Issue
3
Pages
229–235
Publication date
5 February 2026
Language
EN