Comparative Performance of a Routine Immunoturbidimetric Glycated Hemoglobin Assay and High-Pressure Liquid Chromatography Method
Abstract
Introduction Hemoglobin A1c (HbA1c) testing is used for the diagnosis of diabetes and glycemic monitoring. This study compares an immunoturbidimetric assay with high-performance liquid chromatography (HPLC) to assess analytical agreement and determine the clinical suitability of the immunoturbidimetric method for diagnosis versus long-term monitoring in a local setting, in alignment with international standards. Methods A comparative cross-sectional study was conducted in a routine laboratory in Madagascar. Samples were tested on the same day using the Bio-Rad D-10 (Bio-Rad Laboratories, Hercules, CA) and Mindray BS-300 (Mindray, Shenzhen, China) automated analyzers. Analytical accuracy was assessed using control materials. The two methods were compared by evaluating the correlation using Pearson's coefficient (r) and linear regression, and the concordance using the Bland-Altman plot and Passing-Bablok regression. The HbA1c results were divided into a total group and subgroups (HbA1c < 6.5%, 6.5-8%, and > 8%). Results Intra-assay, inter-assay, and overall precision are satisfactory (coefficients of variation (CV)
Fields of Study & Subjects
Diabetes
Cardiovascular Risks
and Lipoproteins
Metabolomics and Mass Spectrometry Studies
Clinical Laboratory Practices and Quality Control