Preparation and Evaluation of a Calcification Phantom for Mammary Gland MRI

Read full text
Authors: Maiko Hashimoto, Norio Hayashi, Masahiko Takahashi
Cited by: 0
References: 22

Abstract

Purpose Microcalcification detection is vital for early breast cancer diagnosis, treatment, and survival; however, no MRI phantom currently exists for parameter optimization. Therefore, this study aims to develop a breast MRI phantom that simulates calcifications. Methods A gelatin-sucrose mixture was prepared, with calcification samples placed on the same plane. MRI using multiple sequences and mammography were performed. Depiction ability was assessed based on contrast, visual evaluation, and statistical analysis. Results Among the 0.75-mm samples, eggshell and glass beads showed similar contrast on X-ray and quantitative susceptibility mapping (QSM). While calcifications ≥0.50 mm were identifiable on X-rays, 0.75-mm calcifications were visible on susceptibility-weighted imaging (SWI), T2-weighted imaging (T2WI), and QSM. Scores obtained for eggshell on T1-weighted imaging (T1WI)/T2WI, chicken bone on T1WI/T2WI, and glass beads on SWI were significantly lower than those obtained on X-ray. Conclusion A microcalcification phantom enabling X-ray-MRI comparison was developed and validated as useful for MRI-based evaluation of calcifications.

Fields of Study & Subjects

MRI in cancer diagnosis Digital Radiography and Breast Imaging Breast Lesions and Carcinomas

Publication details

Journal / source
Cureus
Volume
18
Issue
2
Pages
e103315–e103315
Publication date
9 February 2026
Language
EN