Correction: Whole-Body Dose of Patients Treated With ZAP-X Intracranial Gyroscopic Stereotactic Radiosurgery

Read full text Summarise with VIRAT
Authors: Georg A. Weidlich
Cited by: 0

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

During the delivery of radiosurgery treatments, a small dose is given unintentionally to the areas of the patient’s body outside the primary treatment volume. It is a common practice for radiotherapy platform developers to minimize the dose outside the primary radiation field by providing X-ray target shielding in the beam-defining collimator head of such systems. The amount of such doses was measured for the ZAP-X® delivery platform by designing a phantom-based set-up mimicking the patient and measuring the dose with an ionization chamber and OSLDs. Furthermore, during several clinical ZAP-X radiosurgery treatment deliveries, the dose received by patients was measured with film-based dosimeters. The results were determined as absolute values and as a percentage of the primary dose.

Fields of Study & Subjects

Advanced Radiotherapy Techniques Radiation Therapy and Dosimetry Radiation Dose and Imaging