A new balanced steady state free precession (bSSFP) sequence for the use of Overhauser magnetic resonance imaging at ultralow-field

Summarise with VIRAT
Authors: K. Buckenmaier, Friedemann Bullinger, Pavel Povolni, R. Pohmann, Joern Engelmann, M. Beyerlein, K. Scheffler, L. Kuebler, A. Martins
Cited by: 0

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

Introduction: Ultralow-field MRI (ULF MRI) offers unique clinical potential due to its affordability, accessibility, and intrinsic tissue contrast.[1] To overcome its low signal-to-noise ratio (SNR), Overhauser dynamic nuclear polarization (ODNP) hyperpolarizes the sample using chemical radicals,[2] which also can be functionalized.[3,4] Here, we introduce a novel balanced steady state free precession (bSSFP) based sequence optimized for ODNP, significantly improving sensitivity in ULF MRI applications in vitro and in vivo. Methods: For in vivo experiments, we performed continuous hyperpolarization via ODNP in combination with a SQUID (superconducting quantum interference device)-based ULF MRI system operating at B0 ≈ 750 µT. To acquire images, we implemented a modified bSSFP sequence (bSSFP180) with 180° pulses at TR/4 and 3·TR/4 (Fig. 1). Note that this sequence requires hyperpolarization because the additional inversion pulses suppress thermal magnetization build-up. ODNP excitation at approximately 90 MHz was generated using an 80-cm inner-diameter low-pass birdcage coil, suitable for rats. Carboxy-PROXYL (CP) radicals were injected subcutaneously, intraperitoneal, or intravenously. The concentration and volume of the injected radical solution were adjusted to reach an in vivo concentration of approximately 2 mM. Results/Discussion: Here, we present the first in vivo experiments using continuous hyperpolarization through ODNP in combination with a SQUID-based system. The novel bSSFP180 sequence allowed us to acquire 3D images with a resolution of about 2 mm isotropic in 4 min and 30 s (Fig. 2). Subcutaneous injections enhanced the MR signal near the injection site; however, no detectable signal was observed for intraperitoneal injections. Following intravenous infusion, signal enhancement in the bladder was consistently observed after 10-20 minutes of continuous imaging of both kidneys and the bladder; however, the injection rate varied significantly between experiments. A significant advantage of the bSSFP180 sequence is its ability to prevent banding artifacts, unlike conventional bSSFP sequences used for Overhauser MRI.[5] Additionally, the bSSFP180 spin-echo nature improves robustness against B0 field inhomogeneities and drifts. Conclusion: The bSSFP180 sequence provides significant advantages for ULF MRI with ODNP, particularly in Halbach-based low-field systems where field instability and inhomogeneity limit performance. Using this ULF MRI approach, we obtained the first ODNP-enhanced in vivo images, with CP radicals acting as hyperpolarized contrast agents in the kidneys and bladder. Future radical functionalization could further enable responsive ULF MRI applications.

Fields of Study & Subjects

Atomic and Subatomic Physics Research Advanced MRI Techniques and Applications Advanced NMR Techniques and Applications

Publication details

Journal / source
MPG.PuRe (Max Planck Society)
Publication date
1 January 2026
Language
EN